Home > Product Show

YNi-2316

YNi-2316 Automatic Online Monitor of Nickel in Water

After mixing the water sample with the regulator, the nickel ion reacts with the indicator in the presence of the ammonia environment and the indicator to form a colored complex. The analyzer detects the color change, and converts the change into nick…

Order Online
  • Product Principle
  • Product Features
  • Technical Parameters

After mixing the water sample with the regulator, the nickel ion reacts with the indicator in the presence of the ammonia environment and the indicator to form a colored complex. The analyzer detects the color change, and converts the change into nickel ion value and outputs it. The amount of colored complex is equal to that of nickel ions.

After mixing the water sample with the regulator, the nickel ion reacts with the indicator in the presence of the ammonia environment and the indicator to form a colored complex. The analyzer detects the color change, and converts the change into nickel ion value and outputs it. The amount of colored complex is equal to that of nickel ions.

Characteristics:

Valve components selection. It can select the reagent sampling sequence, with flexible channels, variable functions, minimum dead volume, easy maintenance, long service life and other advantages.

Printing function. It can be connected to a printer (optional order) to print the measured data in time.

It adopts 7-inch touch screen, with simple operation interface, easy for users to learn, operate, and maintain.

Super large data storage capacity. It can realize storage of historical data for no less than 5 years (measurement interval is 1 time/1h), which greatly meets the needs of customers.

Automatic leakage alarm function. When reagent leakage occurs, the instrument will automatically alarm to remind the user to perform maintenance.

Light signal recognition function to ensure high accuracy of quantification.

Simple and convenient chemical reagents maintenance. The reagents can be replaced once a month, which greatly reduces the maintenance of the user.

U disk one-click upgrade. The software update is much easier.

1. Method basis: spectrophotometry.

2. Measuring range: 0.01-15 mg/L.

3. Accuracy: no more than ±10% or no more than ±0.02mg/L.

4. Repeatability: no more than ±5% or no more than ±0.02mg/L.

5. Measurement period: The minimum measurement period is 20 minutes. According to the on-site environment, the color development time can be arbitrarily modified within 5 to 120 minutes.

6. Sampling cycle: time interval (10~9999min arbitrarily adjustable) and whole point measurement mode.

7. Calibration cycle: adjustable at any interval any time from 1 to 99 days.

8. Maintenance cycle: Generally, once a month, each time is about 30 minutes.

9. Reagent consumption: less than 0.50 RMB/sample.

10. Output: 2 RS-232; 1 4-20mA.

11.Environment requirements: indoor with adjustable temperature, recommended temperature +5~28℃; humidity ≤90% (no condensation).

12. Power supply: AC220±10%V, 50±10%Hz, 5A.

Baidu