YCN-2381 Automatic Online Monitor of Cyanide in Water

Add phosphoric acid to the sample, heat and digest it under certain pH conditions. Utilizing the characteristics of stronger complexing ability between metal ions and reagent complexes, and cyanide ions, the complexed cyanide ions are decomposed into …

Order Online
  • Product Principle
  • Product Features
  • Technical Parameters

Add phosphoric acid to the sample, heat and digest it under certain pH conditions. Utilizing the characteristics of stronger complexing ability between metal ions and reagent complexes, and cyanide ions, the complexed cyanide ions are decomposed into cyanide ions, which are distilled out in the form of hydrogen cyanide and absorbed by sodium hydroxide to detect the concentration.

 Characteristics:

Valve components selection. It can select the reagent sampling sequence, with flexible channels, variable functions, minimum dead volume, easy maintenance, long service life and other advantages.

Printing function. It can be connected to a printer (optional order) to print the measured data in time.

It adopts 7-inch touch screen, with simple operation interface, easy for users to learn, operate, and maintain.

Super large data storage capacity. It can realize storage of historical data for no less than 5 years (measurement interval is 1 time/1h), which greatly meets the needs of customers.

Automatic leakage alarm function. When reagent leakage occurs, the instrument will automatically alarm to remind the user to perform maintenance.

Light signal recognition function to ensure high accuracy of quantification.

Simple and convenient chemical reagents maintenance. The reagents can be replaced once a month, which greatly reduces the maintenance of the user.

U disk one-click upgrade. The software update is much easier.

1 Method basis: spectrophotometric detection.

2 Measuring range: 0-0.5 mg/L.

3 Accuracy: no more than ±5% or no more than ±0.005mg/L.

4 Repeatability: no more than ±5% or no more than ±0.0005mg/L.

5 Measurement period: The minimum measurement period is 20 minutes. According to the on-site environment, the color development time can be arbitrarily modified within 5 to 120 minutes.

6 Sampling cycle: time interval (10~9999min arbitrarily adjustable) and hourly measurement mode.

7 Calibration cycle: adjustable at any interval any time from 1 to 99 days.

8 Maintenance cycle: 1 time/month.

9 Reagent consumption: less than 0.1 RMB/sample.

10 Display: 10-inch color screen, 1024*600, 16.77 million true colors.

11 Output: RS-232, 4-20mA (optional).

12 Communication protocol: Modbus RTU, new national standard protocol, dynamic management and control protocol, and others.

13 Environment requirements: indoor with adjustable temperature, recommended temperature +5~28℃; humidity ≤90% (no condensation).

14 Power supply: AC220±10% V, 50±10% Hz, 5A.

 